User Tools

Site Tools


atmos:instruments:wcm:home

This is an old revision of the document!


Problems and Solutions

WCM

Problem (Spring 2025)

Introduction

During previous field campaigns in which the WCM 3000 was used it had been noticed that the initial value during no water content present was a non zero value. This is still noticed in the WCM 3000 where on initial start up the probe indicated that there is water content present. This issue was normally resolved with post processing of the data from past campaigns where the non zero data would be adjusted down to zero and then further studied after that correction was applied.

Cause

In the fml.300 table that is present on the m300 computer and is used by the WCM data collection software it was noticed that the code that is supposed to correct for this non zero value wasn't included in the table. There was however correction code in there but it appeared to be from the older WCM model in which it had a compensation term that would remove this non zero value and instead keep the probe at a zero value until water content is present. '

Solution

In the manual for the WCM 3000 probe there is a section in which a couple different was are discussed as to how to get the probe to go to a zero value during the non presence of water. (These will be added here).

Problem (Spring 2025)

Introduction

During the use of the WCM it was noticed that when compared to other instruments such as the Cloud Droplet Probe, RICE probe, and the hot wire boom probe there is a decay present after the water content should have gone to zero over the probe head. This is now something that is being lab tested to better understand what is going on with the probe and to try and correct this invalid data that is being collected. Below are references to where the data is located for this particular lab run.

  • Reference: /nas/und/NorthEast/2025/Lab/CH423/GroundData/20250416_165626/PostProcessing
  • Reference: /nas/und/NorthEast/2025/Lab/CH423/GroundData/20250416_165626/Camera_Data

Below is a side by side comparison of what was recorded during this test by both the probe head and the video which in the video shows when water droplets are present and when they aren't. The graph is also marked as to roughly where the video is showing data at and where the corresponding area is on the graph.

Video Comparison

NOTICE

One thing to note with the data displayed in the graphs is that the zero baseline is at 2.2 instead of being at zero. This is another noticeable issue with the M-300 and the way it records data. I will refer to this baseline as 'zero' in the rest of the analysis.

Analysis

The delay between when the WCM displays a 'zero' value and when the probe should be at 'zero' is around about a five or so second delay which can be observed by the graphs. Which after applying a 66.7% data validity factor to the delayed seconds this comes out to being about a 3.34 second delay where data is being displayed that shouldn't be. This is of course just one case but, during this particular data set I did a few other runs where I would run a minuet without water and then a minuet with water. Below are the other runs that I did which all show this similar decay that occurs right after the water is cut off from flowing over the probe head.

Decay Graphs

Top left - 17_02_00 Graph, Top right - 17_04_00 Graph, Bottom left - 17_06_00 Graph, Bottom right - 17_08_00 Graph

Each graph is marked at roughly where the video indicated no more water being present over the probe head.

The way that I determined the delay interval was by determining that each tick on the graph is roughly one second of data and then counting the ticks it took to get to a baseline value. The way I determined this baseline was by observing where the data mostly stabilized at so this is mostly subjective in terms of determining this value. But I mostly settled on a value of at or below 2.2 on the x-axis. The corresponding values that I then found for each graph are as follows; 6 seconds for 17_02_00, 6 seconds for 17_04_00, 5 seconds for 17_06_00, and 7 seconds for 17_08_00. This then came out to be an average of 6 seconds of data present during this decay. Then after applying the 66.7% validity to this average I found that there is about 4.00 seconds of incorrect data during these runs.

Suggested Solutions

After contacting SEA they suggested that a possible reason for this issue is the PID controller that is in the WCM may not be set to correct settings thus resulting in a decay in the graph after water content becomes non present. It was also shared that there are a couple different settings that they tested when calibrating the WCM 3000 where they gave us what the default values should be along with a mid aggressive and aggressive setting for the PID board.

Implementation

These values were all tried and each one preformed in a manner that wasn't necessarily similar to what SEA had gotten for results after using these values. The output was more so aggressive when set to an aggressive setting as it should be but there were still noticeable decays in the graph even with these more aggressive values.

From here values were changed based upon the way that PID controllers should be calibrated where you first start with the Proportional term and get it to an appropriate value that causes the instrument to operate stable and then you adjust the Integral and Derivative terms as needed to get the output from the instrument to be as stable and responsive as possible. After changing these values many times there was still the same decay issue present in the recorded graphs.

Solution One to Decay

Checking the fml.300 table within the m300 computer it was noticed that the data this is being recorded is being averaged which seemed like it could be causing an issue with our recorded data and thus was removed to see the effect. After removing this averaging function there was a noticeable improvement in the data collected where now instead of these long 5 to 7 second data delays there was now only about a 2 to 3 second delay present with the data (This delay can now be fixed by tuning the PID controller).

Solution Two to Decay (Tentative)

Now that the long decay was removed from the graphs by removing the averaging function in the fml.300 table the PID controller should now have a more noticeable impact on the data collected. Another thing to note which was found during the initial PID tuning was that the derivative term will also contribute to a decay since the derivative term is acting as a dampener to stop the probe from immediately going to zero after water content is stopped. So now taking that into account when tuning the PID board it is known that the derivative term should be set fairly low so as to stop the probe from over dampening itself during data collection. The values that were found that allowed the probe to work fairly well are as follows; (TWC terms Kp- 2000 Ki- 90 Kd- 20) (LWC terms Kp- 9500 Ki- 95 Kd- 20). These values were only tested during no airflow to try and calibrate the probe to be as stead as possible. This was done as a control just to factor out the unpredictability of the airflow that would normally be present over the probe and to make the tuning of the probe more easier.

These above PID values still need to be tested in lab to validate that they are actually good values that should be used during future field campaigns with this probe.

Problem (Winter 2024)

During the setup of the WCM, we were having issues with the power not being supplied to the WCM Power Box which prompted the WCM software to say that ser17 can't open.

Cause

The power connection of the leads from the WCM Power box to the D/C Power supply was not making a good connection with the prongs themselves on the D/C Power supply.

Solution

The leads coming from the WCM power box where squeezed together to help in making better contact with the prongs on the D/C Power supply.

atmos/instruments/wcm/home.1748012691.txt.gz · Last modified: 2025/05/23 15:04 by 127.0.0.1