Line 22: Line 22:
  
 2026/07/24 While attempting to start the radar, we had the same issue of the door switch holding capacitance, and in an attempt to fix it the switch was grounded to the floor of the cabinet. The power supply fuse then blew; it is unclear if this was caused by grounding the door switch or if it was a coincidence. New fuses have been ordered and are on the way.  2026/07/24 While attempting to start the radar, we had the same issue of the door switch holding capacitance, and in an attempt to fix it the switch was grounded to the floor of the cabinet. The power supply fuse then blew; it is unclear if this was caused by grounding the door switch or if it was a coincidence. New fuses have been ordered and are on the way. 
 +
 +2026/07/30 The door switch signal was changed from critical to normal in the IRIS RCP8 setup. This allows IRIS to ignore the red flag warning that the interlock is open. It was discovered that the transmitter cabinet has it's own interlock detection system that requires the door to be closed long enough to build up voltage, this seems to be working so it is assumed that the short is somewhere between the cabinet and the RCP. The system was ran with the ignored interlock warning and ran for several minutes, detecting the storm west of UND. The program was ended by a shutdown of the RVP8. The RVP8 was repeatedly rebooting, but the DSP socket was not connecting to IRIS. IRIS was able to launch so it can see and interact with the RVP8 computer but not use it's DSP. The RVP8 was left manually turned off and will need to be turned back on on a cooler day. Current temperature wis 86 F with no sun. 
 +
 +2026/7/31 Several attempts were made to get the RVP8 running again this morning. After the initial power on the RVP was pingable from radar1 but IRIS still reported no DSP connection. After running  several test codes while sshed into RVP it hit an error that it could not interact with the IO card. The RVP was then rebooted but now the IFD lights refused to turn on. The RAM was reseated again, and all cable connections were checked. The RVP still is unable to boot, and connected monitor showed no input. The RVP is currently powered off and another attempt will be made on Monday to start the RVP8.
 +
 +2026/8/15 The new Jump Station was installed by UND IT allowing ethernet connection and faster data transfer. It also allows a more secure connection to outside computers.