Tripod Instrumentation Meteorological Measurement System (TIMMS) Online Lab Book and Field Notes
Latest information is given at the top. Times and dates are UTC.
2025/10/22 01:00 UTC\ David Delene: Visit the Glacial Ridge site and the power on the TIMMS SN 00300 was off. The outlet CFI break had tipped for the second time. Turn on two breakers in the Breaker box and the other outside outlets were activated. Moved the instrument extension cord to one of these outlets, which are not CFI outlets by have good covers. Needed to connect via Anydesk and get the Visibility Program to use a different USB device, need to get the USB devices configured correctly.
2025/08/11 19:00 UTC
Bryce R. Rickbeil: Completely unhooked the second temperature and relative humidity sensor from the TIMMS00200 tripod. This was done in response to the data coming from that sensor being very incorrect and very sporadic when compared to the other temperature and relative humidity sensor which was stable and in agreement with other temperature and humidity reports around the Grand Forks area. This probe is still attached to the tripod and just the wiring has been disconnected so no power is going to the instrument any more.
2024/05/08 21:00 UTC
Michael Willette: Restarted IMet-XQ2 code after instrument was not reporting data. Suspect bad data point or no data point issue.
2024/03/14 23:15 UTC
David Delene: Manually started the visibility script since only this instrument was not reporting data in Grafana.