atmos:datacollections:2022_03_15
March 15 2022
* Run 05: SMPS_20220315_Run05
- No notes on this run.
* Run 06: SMPS_20220315_Run06
- Start Time: 14:24 UTC
- CPC Starting Concentration: 6300 #/cm3
- APS Starting Pattern: Flat. Minor peaks around 3-4 μm at 4 #/cm3
- CPC continuing to linearly drop off
* Run 07: SMPS_20220315_Run07
- Start Time: 14:59 UTC
- CPC Starting Concentration: 2800 #/cm3
- APS Starting Pattern: Relatively Flat
- CPC Concentration after 5 min: 2000 #/cm3
- Any minor APS peaks are around 5 #/cm3
- CPC showing asymptotic behavior, signs of flattening out.
- Tapping valves causes spikes in CPC concentrations.
* Run 08: SMPS_20220315_Run08
- Start Time: 15:28 UTC
- Filter Off: 16:12 UTC
- CPC pattern very different. CPC concentrations increased and now relatively stable. Different solution producing different results. CPC concentration stabilized around 380 #/cm3.
- APS Pattern: Flat
- After peak, seeing slight decay in concentration.
* Run 09: SMPS_20220315_Run09 (Acetone Run)
- Start Time: 16:25 UTC
- Purge of Generator: 16:29 UTC
- Peak of purge at 100 #/cm3. Falling off quickly.
- Second peak observed at 16:37 UTC. Unknown what the cause is.
- Filter Start Time: 16:41 UTC
- Back to background levels after purge at 16:44 UTC.
- Train went by at 16:51 UTC. No response in data.
- Tapping valves at 16:54 UTC. Minor peak of 40 #/cm3.
* Run 10: SMPS_20220315_Run10 (Propane Run)
- Start Time: 18:08 UTC
- Filter On: 18:09 UTC
- CPC just background values ( ≥ 0.5 #/cm3).
- APS has very few larger particles. Peak around 0.6 μm at ∼0.2 #/cm3.
- 10 mins in, everything remains the same. APS peaking between 0.6-0.7 μm.
* Run 11: SMPS_20220315_Run11 (IPC 2021 Old Run)
- Start Time: 19:01 UTC
- CPC Concentration Peak: ∼440 #/cm3
- APS Pattern: Flat, no real peaks
- CPC showing linear decrease. ∼350 #/cm3 at 5 mins in.
* Run 12: SMPS_20220315_Run12 (New Filtration System: Duplicate to Run 1)
- Filtration System added: 0.2 #/cm3 background
- CPC hovering around 20,000 with valves open. Counts averaged around 17,000 #/cm3. Background is 0.1-0.3 #/cm3.
- Background 1 to Atmos. 15,000 ratio
- CPC Start Time: 20:05 UTC
- Sample Start Time: 20:30 UTC
- 1 minute sample outside (air without dilution)
- 1 minute sample with dilution
- Repeat process for two of each.
- Generator CPC peak at 1,100 #/cm3. Dropping linearly to 750 #/cm3 3 minutes later.
- CPC holding steady between 700-750 #/cm3.
* Run 13: SMPS_20220315_Run13 (WMI Formula)
- CPC Start Time: 20:55 UTC
- Initial Concentration: ∼20 #/cm3
- Ratio Tests:
- 13-15 background to 40,000 (1 valve)
- 13-15 background to 80,000+ (all valves)
- Concentrations not dipping down to previous background levels. Some contamination possible. Another bypass might be needed.
- Running propane to clear 8“ tubing. No real response. Still 8-10 #/cm3. Needs to be around 1 #/cm3.
- More ratio tests:
- 3-6 background to 14,000-15,000
- 5-7 background to 12,000-13,000
- 5-7 background to erroneous values.
- Sample Start Time: 22:10 UTC (same as filter start time)
- CPC Concentration: 3,720 #/cm3
- APS pattern: Continued relatively flat pattern ∼ 4 #/cm3
- Fans once again died at the end. (overheated)
atmos/datacollections/2022_03_15.txt · Last modified: 2022/11/16 20:08 by 127.0.0.1