Vaisala HMK15 Humidity Calibrator

Parts

  • Main Board
    • Salt Solution Chambers (x2)
    • Sensor Cap Attachment (x2)
    • Sensor Cap Plugs (x10)
    • Thermometer
  • Measuring Cup
  • Plastic Probe/Stirring Stick
  • Adapter Rings (x3)
  • Salts
    • NaCl (75% RH)
    • K2SO4 (97% RH)
  • Calibration/Preparation Date Tags
  • Certificate of Calibration/Inspection

Calibration Process

Need:

  • HMK15 Humidity Calibrator
  • Vaisala Humidity Sensor
  • Deionized water

Steps:

  1. Prepare salt solutions
    1. gg
    2. Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared.
  2. Wait 24 hours
  3. Calibrate Instrument