atmos:citation:instruments:uhsas:home
Table of Contents
Ultra High Sensitivity Aerosol Spectrometer
- Description:
- Manufacturer: Droplet Measurement Technologies
- Serial Number:
- Measured Particle Size Range: 60 to 1000 nm
Notes
- On March 23, 2011 the reference voltage for the UHSAS was about 2.2 V. If this voltage falls below 2.0 V the instrument is no longer functioning properly and the measurements are unreliable. On March 24, the reference voltage was about 2.4 V.
Calibration
- Calibration was done at UND on March 24, 2011