atmos:citation:instruments:timecodegen:home
Table of Contents
Time Code Generator
Battery Change
- Time Code generator internal battery was changed on 07/25/13
- Battery is a Lithium Coin 3V Part # CR2430
- Extra Battery is located in the Airport Lab in the Instrument Equipment cabinet in the bin labeled “EXTRA”
- Battery change directions:
- Disconnect the Comm port, Ant, Power, and Cannon Plug
- Remove the four screws on the front plate (four corners)
- Slide the front plate straight out !Caution the board it connected to the plate DO NOT BEND!
- If unable to pull straight out you may need to remove the Time Code Generator from the research rack.
- Replace the battery
- Reverse order to reinstall
Problem
- Time Code generator was outputting data in the wrong time code generator.
Cause
- Time Code generator was in the wrong time zone.
Solution
- Need to use a hperterminal client. Interface settings are 19200, 8 bits, No parity, 1 stop bit, and no flow control. Should receive data every second if connect correctly. You may also need a Null Modem
- Send the code: +00 to set a no time zone offset. Check that the time zone offset is correct by sending S. The S (Status) should return “S+00112” The 00 indicates no time zone offset. If the return is “T” the last 6 digits before that lat/long will show the time. Make sure that time is the current UTC time.