Cloud Spectrometer and Impactor (CSI)

Description

  • Measures total water and droplet size distribution.

Mounting Location

  • External

Air Sampling Location

  • External

Instrument Requirements

  • Power: 28 VDC 10 Amps [Pump] 110 VAC [Source: Paul LeHardy]
  • Weight: 41 Kg (90 lbs) [Source: Paul LeHardy]
  • CSI IPHEX and OLYMPEX projects
Probe Power Amps Items
28VDC #1 3.1 DC-DC converters TDL, CDP
Aux PWR 2.0 TDL, vaporizer, flow solenoid
28VDC #3 3.5 Tip, Fin, Pitot heater
28VDC #2 11.4 All CDP heaters
DC Pump PWR Amps
28VDC 4.0

[Source mikhail.pekour@pnnl.gov Date:01/21/2015]

CSI Pump

  • Manufacturer: Neuberger Inc.
  • Type: MPU 1798-N838.0-9.05
  • Serial # : 1/857776
  • Power requirements: 24 V DC 4 amps
  • Sent from DMT
  • Manufacturer:Neuberger Inc.
  • Type: MPU 1798-N838.0-9.05
  • Serial # : 1/1251252
  • Power requirements: 24 V DC 4 amps
  • Sent from ARM
  • Note: Additional control box attached

Data Collection

Picture

Network Connection

  • IP Address: 192.168.0.140.0
  • Subnet: 255.255.255.0
  • Gateway: 192.168.0.1
  • Connect to Network Router and attach laptop to network that uses a different address.
  • VNC password is: g1
  • Connect with the CSI test cable:
    1. Plug network cable directly to the laptop.
    2. Connect an Ethernet crossover plug to the test cable prior to connecting to the network hub.
  • Citation Network
    • Make sure the “Uplink” port is disconnected on the network box (CSI will not connect to the network if that port has a cable in it)

RS-422 Connection

  • Used to output data to another computer and to remotely control the CSI flows.
  • Communication is as 9600 baud 8, N, 1, with no flow control.
  • Data will only be sent when the unit is operating under flight conditions.
  • Data will be sent a 1 Hz intervals.

Laptop connection to CSI Computer

  • Used for data download from CSI computer.
  • Need:
    • “remote” computer (e.g. laptop)
    • crossover network cable
  • Notes:
    • These instructions are from the CSI manual. They haven't been tested yet as of 19 Feb. 2014.
    • The manual states the remote computer should be running Windows 2000 or XP.
  1. Remove the cover plate from the left side of CSI dome.
  2. Connect the monitor crossover cable between the impactor and the remote computer. Power up the remote computer.
  3. Power up the impactor.
  4. On the remote computer, click on Start, right click on My Network Places, click on Search for Computers, enter 10.10.10.10 and hit Search. The system should find the impactor directory C.
  5. Double click on C, and the directory listing of the impactor will appear. Go to the data directories and copy over the data as needed. It is recommended that the data be deleted from the impactor computer after it has been archived at another location.

Output Data Channels

Main Data File

Short Name Units Long Name
time HH:MM:SS Time
meas_supp lpm Measured Supply
meas_bypass lpm Measured Bypass
meas_sample lpm Measured Sample
count_flo lpm Counter Flow
tdl_h2o g/m3 TDL H2O
tdl_press mb TDL Chamber Pressure
tdl_temp C TDL Cell Temp
tip_temp C Tip Temp
tube_temp C Tube Temp
amb_temp C Ambient Temp
rtd_resist Ohms RTD Resistance
field_wv Field Calibrated Water Vapor
twc mg/m3 Total Water Content
air_speed m/s Air Speed
temp_5 Temp 5
temp_6 Temp 6
temp_7 Temp 7
volt_6 Volt 6
volt_7 Volt 7
numb_conc #/cm3 Number Concentration
vol_conc um/cm3 Volume Concentration
lwc g/m3 Liquid Water Concentration
mvd um/cm3 Median Volume Diameter(MVD)
eff_dia um Effective Diameter

TDL Output File

Short Name Units Long Name
time HH:MM:SS Time
tdl_1 g/m3 TDL 1
tdl_2 mb TDL 2
tdl_3 C TDL 3
pp2f PP2F
power Power
index Index
zero Zero
t_laser T laser
t_sink T sink

Calibration Coefficients

Cal.Coeff. Tab: The following can be changed using this formula. Scaled Value =(slope*raw value)+offset

  • 1. Static slope
  • 2. Dynamic slope
  • 3. Supply slope
  • 4. Bypass slope
  • 5. Sample slope
  • 6. Static offset
  • 7. Dynamic offset
  • 8. Supply offset
  • 9. Bypass offset
  • 10. Sample offset

Current Settings:

  • 1. Static slope: 243.16
  • 2. Dynamic slope: 80.84
  • 3. Supply slope: 1.0038
  • 4. Bypass slope: 0.9098
  • 5. Sample slope: 0.9012
  • 6. Static offset: -231.71
  • 7. Dynamic offset: -67.071
  • 8. Supply offset: 0.0392
  • 9. Bypass offset: 0.1355
  • 10. Sample offset: 0.0415

Raw tab TDL Cal. Coef: C_0:1 C_1:0 C_2:0

Scaled Values are in the Raw tab and in the data file

  • 1. Static pressure(mb): channel 34
  • 2. Dynamic pressure(mb): channel 35
  • 3. Supply set: channel 24
  • 4. Bypass set: channel 25
  • 5. Sample set: channel 26
  • 6. Static pressure raw(V): channel 37
  • 7. Dynamic pressure raw(V): channel 38
  • 8. Supply flow raw(V): channel 39
  • 9. Bypass flow raw(V): channel 40