atmos:citation:instruments:cdp:home
Table of Contents
Cloud Droplet Probe (CDP)
- Description: Cloud Droplet Probe
- Manufacturer: Droplet Measurement Technologies
- Model: CDP2
- Serial Number: 0901-48
- Weight (UND has short version): 3.5 lb
- (Newer long version): 3.0 lb
- Size (UND has short version): 8.5“ (long), 6.0” (wide), 4.5“ (height)
- Long (larger version) 10.5” (long), 6.0“ (wide), 8.5” (height)
- UND Major Equipment Property Tag Number: 105883
Cloud Droplet Probe Pelican Box Contents
As of 5/16/2021:
- Cloud Droplet Probe (SN: 0901-48)
- Electronics Box (SN: 0901-048)
- Control Board (ABD-0234)
- CDP Probe to Electronics Box Cable (New)
- CDP Probe to Electronics Box Cable (Old)
- CDP Control Box Mounting Plate Hardware
- Electronics Box to Data Acquisition Cable
- RS422- M300 Converter Cable
- Adapter Plate with Screws
- Test Cable with Gender Changer
- P4 Cable End and Cap (Within a bag)
- Screws and Miscellaneous Equipment (Within a bag)
DMT Calibration Files
-
- Firmware: 169D installed
-
- Sealed Windows,tuned laser heater to 25 C, installed PBP-Dynamic Threshold Firmware - Corrected 72 Board loading, installed 80 um pinhole. The serial interface was set to RS422.
- Spreadsheet of bead performance check results is located in the /nas_remote/PerformanceChecks/CloudDropletProbe directory.
Laptop Connection Configuration and Software
- RS-422
- Baud Rate: 57600
Citation wiring is configured to directly connect to laptop using RS422 to RS232 converter. If a 422 to USB converter is used then an adaptor is necessary. Furthermore, an adaptor is needed to connect to the M300 Serial box.
The DMT PADS 3 software can be used to connect and test the CDP. When installing the PADS 3 software it is important to copy the CDP directory to C:/Program Files/PADS 3. The PADS 3 software is locate at /nas/Instrumentation/CloudDropletProbe/Software.
Problems and Solutions