User Tools

Site Tools


atmos:citation:field:impacts2021:home

NASA IMPACTS 2022

  • Project: IMPACTS - Investigation of Microphysics and Precipitation for Atlantic Coast-Threatening Snowstorms
  • Dates: January and February 2022
  • Location: North East (United States)
  • Platform: NASA P3
  • Tail Number: N426NA
  • Flight Level Processing Script: process_all_impacts

Links

Processing

Analysis

Documentation

Note: The files below may be updated or added to frequently. Please view the most recent version on the UND “Aircraft” server in the locations listed.

  • Aircraft Configuration Document: /nas/und/NorthEast/2022/Aircraft/P3_N426NA/Documents/Configurations
  • Probe Operator Checklists: /nas/und/NorthEast/2022/Aircraft/P3_N426NA/Documents/Checklists
  • Probe Maintenance Checklists: /nas/und/NorthEast/2022/Aircraft/P3_N426NA/Documents/Checklists
  • M300 Tables: /nas/und/NorthEast/2022/Aircraft/P3_N426NA/Documents/M300_Table_Archive
  • Ground Test Data: /nas/und/NorthEast/2022/Aircraft/P3_N426NA/GroundData
  • Flight Data: /nas/und/NorthEast/2022/Aircraft/P3_N426NA/FlightData

Items to Bring for January 2022

  • Probe Covers (HVPS, CDP, King, etc).
  • IPAD and holders
  • Cable Ties
  • Spare Network cables
  • Spare USB to Network adapters
  • Spare label maker tape
  • USB sticks for instruments

Items and Issues to Address Post-field Project

  • Set the HVPS3-B computer to not boot on power on. Have to remove cover and set jumper switch.
  • See about getting old M300 system booted or updated with new system.
  • Holder for in front of 2DS for mist generator.

2022/01/11 Flight Items

  • 2DS Can Low Pressure
    • Issue - The 2DS pressure went < 9 psi; while the HVPS3 pressures remained high.
    • Cause - Like cause is a bad o-ring or seal on Can. Will remove probe and check o-ring.
  • Downlinking Data
    • Issue - On NASA system not getting down link packages from the M300. Will ground trouble shoot.
  • Hawkeye TimeSync
    • Issue - Have to manually set the time on the Hawkeye computer.
    • Cause - Works on other computers and can ping the NTP server. Will ground trouble shoot.
    • Fix - Needed to adjust the Time server IP address in the 'NTP Time Server Monitor (the application is on the desktop). Had to click on the NTP Configuration File tab and scrolled down to replace the old time server IP address. Next, we had to run the 'SPEC NTP Sync' application (also located on the desktop) which allowed the time to sync up with the P-3. Now the time should sync up with the aircraft upon boot-up.
  • M300 WCM Data
    • Issue - Need to adjust M300 configuration so Data status window correctly displays getting WCM data.

2022/01/17 Flight Items

  • HVPS-3A Connection issue
    • Issue - Software was not able to connect the the probe. Was able to ping the probe through the software and through a command prompt. Still no data.
    • Cause -
    • Fix -

Getting Ready for 2023 Deployment

Items to Buy

Item Purpose of Purchase Link
UPS To be installed on racks to help power distribution to computers and instruments https://www.amazon.com/Tripp-Lite-Line-Interactive-Rackmount-SMART500RT1U/dp/B0002CUA6K/ref=psdc_764572_t1_B0785MRNVZ?th=1
Dual Monitor KVM Link to be provided
Tool Box To replace old tool box that is broken https://www.lowes.com/pd/CRAFTSMAN-Portable-Tool-Box-20-5-in-Ball-bearing-3-Drawer-Red-Steel-Lockable-Tool-Box/1001107526
Wireless Keyboard - Right Rack Installing keyboard on IMPACTS rack https://www.walmart.com/ip/VicTsing-Wireless-Touchpad-Keyboard-Ultra-slim-All-in-One-Wireless-2-4Ghz-Keyboard-with-Built-in-Multi-Touch-Touchpad-Black/271962916?wmlspartner=wlpa&selectedSellerId=101066168&&adid=22222222227460380764&wl0=&wl1=g&wl2=c&wl3=538800917884&wl4=pla-1400514077224&wl5=1021449&wl6=&wl7=&wl8=&wl9=pla&wl10=377878558&wl11=online&wl12=271962916&veh=sem&gclid=CjwKCAjwzeqVBhAoEiwAOrEmzdPKYwhYgpfvQzATkkNYb1CttarQDT7nUk4ip_vAscLZwsdLXRnHthoCq2UQAvD_BwE&gclsrc=aw.ds
Sniffer Fluid For leak detection on instruments Link to be provided
Reusable cable ties To tie wires for impacts racks https://www.amazon.com/dp/B09ST7JMWC/ref=sspa_dk_detail_5?pd_rd_i=B09ST7JMWC&pd_rd_w=y3YT3&content-id=amzn1.sym.3481f441-61ac-4028-9c1a-7f9ce8ec50c5&pf_rd_p=3481f441-61ac-4028-9c1a-7f9ce8ec50c5&pf_rd_r=Q4C90HPBCE9S0QF2FTG1&pd_rd_wg=ZUTE0&pd_rd_r=2ca247a7-145b-4918-b19c-96b3c9794438&s=hi&spLa=ZW5jcnlwdGVkUXVhbGlmaWVyPUEzNDg2NkdXVE8yOUNNJmVuY3J5cHRlZElkPUEwNDAwMzI3MUlRMEVLUVBSODhZRiZlbmNyeXB0ZWRBZElkPUEwNTY0NDMwMVVORFpZSDZGTVRNOSZ3aWRnZXROYW1lPXNwX2RldGFpbF90aGVtYXRpYyZhY3Rpb249Y2xpY2tSZWRpcmVjdCZkb05vdExvZ0NsaWNrPXRydWU&th=1
Wood Working Clamps To hold down instruments while being tested on bench Link to be provided
Motherboard expansion card clips For M300 to keep serial cards in place Link to be provided
Cleaning Swabs To clean instruments https://www.amazon.com/Swabs-4-25-Cleaning-Electronic-Devices/dp/B004Q2JH0C/ref=sr_1_1_sspa?dchild=1&keywords=chemtronics+swab+and+applicator&qid=1631807129&s=beauty&sr=1-1-spons&psc=1&spLa=ZW5jcnlwdGVkUXVhbGlmaWVyPUEyVUVWQ0JaSjA3QUE0JmVuY3J5cHRlZElkPUEwNzEzMDYxMllPSUtXNTQ4OFVJMyZlbmNyeXB0ZWRBZElkPUEwMTY2NzE2M0NRQ1c2TEU5UTYwWiZ3aWRnZXROYW1lPXNwX2F0ZiZhY3Rpb249Y2xpY2tSZWRpcmVjdCZkb05vdExvZ0NsaWNrPXRydWU=
Aircraft Hexshaped Bolts - AN3-5A (Quantity -100) Need more bolts for mounting on racks https://www.aircraftspruce.com/catalog/hapages/an3.php?clickkey=4917239
Aircraft Spring Nut - #10/32 thread (Quantity - 40) Need more spring nuts to mount on racks https://www.grainger.com/product/GRAINGER-APPROVED-Spring-Nut-G-Style-Spring-4CUG4?opr=PDPRRDSP&analytics=dsrrItems_4CUG7

Status of Items Purchased

Item Date Item was Ordered Date Item was Received Date Item was Installed Items Current Location (Date)
UPS 7/8/22 7/15/22 Clifford Hall Lab
Cleaning Swabs 7/8/22

Things to Be Addressed

atmos/citation/field/impacts2021/home.txt · Last modified: 2023/11/04 06:32 by 127.0.0.1