Table of Contents

Table Misconfiguratiton

Date

Problem was on the March 26, 2010 flight.

Problem

The M300 tables were configured to only record type 53, 6, 10, 11 data from the HVPS. The time information (type 7) was not recorded. This makes processing impossible since we don't have time information for the images.

Cause

Using untested tables.

Solution

Review of the M300 test directory test shows that type 7 data is not part of the HVPS configuration and hence the processing had to be modified to not have to use type 7.