How to Remove CPI Probe Case

Procedure

  1. Remove 14 total hexagonal screws along the canister.
  2. Remove bottom of the casing.
  3. Remove 3 Philips screws along the side of the rectangular casing.
  4. Unscrew the hex-head post on the bottom of the instrument by the circuit board box.
  5. Pull instrument free from casing.

View of 7 of the hexagonal screws along the bottom of the canister.

Figure 1: Picture of the bottom of the CPI casing with yellow arrows to show location of hexagonal screws along the container.

View of 3 Philips screws along the side of the rectangular casing.

Figure 2: Picture of the top of the CPI container with yellow arrows to show the 3 Philip-head screws

 View of the hex-head post.

Figure 3: Picture of the hex-head post located by the circuit board box.