Note #1 - January/February 2011 Modifications at DMT. Information obtained by talking with Gary Granger on the phone (Feb 14, 2011). The Firmware was updated to the latest version which should solve any sample volume change issue. The pin hole was changed from not having one (~2.5 - 3.0 mm size) to a 800 um pin hole. The Threshold tables are different. Gary Granger sent new values via email to David Delene. The minimum value is still at 20. The sample area that you should use is still 0.24 mm^2.
Date | Location | Notes/Reason for Relocation |
---|---|---|
July 03, 2012 | Citation, right pylon | Received CDP back from DMT on July 2nd. (CDP S/N 0901-048) |
May 29, 2012 | DMT | Sent in to DMT for new board*. |
February 16, 2011 | Received Probe from DMT | See Notes #1 above for changes. |
January 5, 2011 | DMT-RMA#1360,S/N0901-048 | Sent in to fix 3 issues w/ probe. Pin hole sizing, Firmware update, RS232 to RS 422, and pacs software. |
February 14, 2010 | Citation Research Aircraft | Installed on Citation Research Aircraft. |
February 10, 2010 | Atmos. Airport Lab | Brought probe from Clifford Hall to install on Citation. |
February 10, 2010 | Atmos. Department | Received the probe back from SEA. |
January 14, 2010 | SEA | Sent probe to SEA so they could develop particle-by-particle software. |
January 7, 2010 | Rm 423 Clifford Hall, UND | Changed number of channels from 20 to 30 |
December 21, 2009 | Rm 423 Clifford Hall, UND | Received back from DMT, with short 3“ stut instead of long stut |
Sept. 3, 2009 | Rm 423 Clifford Hall, UND | |
August 27, 2009 | Grand Forks, ND Airport | |
August 24, 2009 | Rm 423 Clifford Hall, UND | Calibration Check |
August 13, 2009 | Grand Forks, ND Airport | To have adapter made so it can be attached to the citation |
August 5, 2009 | Rm 423 Clifford Hall, UND | Received back from DMT |
Mid June 2009 | DMT | Sent in to fix sizing issue* |