2D-C Spinning Disk Calibration

Calibration

  1. Put the bracket around the 2D-C arms and attach the spinning disk to the bracket. Make sure the red dots on the bracket and spinning disk are side by side.
  2. Align the disk so that the center of the big dot is right in the middle of the 2D-C laser.
  3. Turn on the spinning disk.
  4. Set the true air speed for the probe to 25 m/s.