===== Setting up the Veinasa Visibility Sensor ===== Manual: [[|Veinasa Visibility Sensor Manual]] === Parts === The Vaisala temperature and humidity probe has a few key components when it comes to setting up the instrument. *Main Body {{:atmos:citation:instruments:setup_veinasa_visibility_sensor:screenshot_2026-07-14_154844.png?200|}} *Wires {{:atmos:citation:instruments:setup_veinasa_visibility_sensor:screenshot_2026-07-23_172203.png?200|}} {{:atmos:citation:instruments:setup_veinasa_visibility_sensor:screenshot_2026-07-23_172222.png?200|}} {{:atmos:citation:instruments:setup_veinasa_visibility_sensor:screenshot_2026-07-23_172239.png?200|}} * Green - T/R- * Yellow - T/R+ * Red - Power (positive) * Black - Grounding power (negative) === Setting up the Sensor === By now, you should have your Raspberry Pi computer completely set up, with a DAQ HAT. To continue, you will need: * some sort of DC Power Jack connector {{:atmos:citation:instruments:setup_vaisala_temperature_and_humidity_probe:screenshot_2025-11-24_203842.png?100|}} * USB to RS422/485 converter * Extra wire - Connect the DC Power Jack to power - Connect the red wire to the positive connection - connect the black wire to the negative connection - Wire the probe into the Converter - Connect the yellow wire to T/R+ - Connect the green wire to T/R- - Connect the extra wire to GND and the negative connection on the DC power jack, as shown above. - If you have properly set up the ADPAA, you can now run the python program that coincides with your instrument and observe the data.