====== Cloud Imaging Probe ====== *Description: Cloud Imaging Probe *Manufacturer: Droplet Measurement Technologies *Serial Number: *Measured Particle Size Range: 25 um - 1.55 mm *Owner:Center for Interdisciplinary Remotely-Piloted Aircraft Studies (CIRPAS) ===== Problems and Solutions ===== ==== Problem ==== The CIP was showing housekeeping data as well as mean particle diameters and concentrations, but no images. ==== Cause ==== Bad connection between the CAPS card and the M300 computer. ==== Solution ==== Reset the CAPS card in the M300 computer. ======Calibration Constant====== ==== From the July 16, 2011 flight=== 67600 to 68300 Seconds from Midnight ===Nose Pitot vs IAS=== ==Slope: 2.231 Offset: 0.00== ==Slope: 1.163 Offset: 42.939== {{:atmos:citation:instruments:cip:nose_pitot_cplot.png|}} ===Wing Pitot vs. IAS=== ==Slope: 2.2504 Offset: 0.00== ==Slope: 1.3284 Offset:37.09== {{:atmos:citation:instruments:cip:cplot.png|}}