===problem (Summer 2025)=== | ===problem (Summer 2025)=== | ||
- | Mass concentration wasn't being recorded in from the PM-10 size bin. | + | Mass concentration wasn't being recorded in from the PM-10 size bin. Then later PM-2.5 was found to be giving inaccurate readings. |
===Cause=== | ===Cause=== | ||
- | There are two magnets | + | One possible cause was the fact that the magnets weren' |
+ | |||
+ | The other possible and most likely cause is that the bottom power board in the TEOM computer housing cabinet was completely corroded | ||
===Solution=== | ===Solution=== | ||
- | The way that Thermo Scientific recommended to fix this issue was by replacing | + | To fix the issue with the PM-2.5 not giving accurate readings |
+ | |||
+ | To fix the issue with the PM-10 not being read in the most plausible solution would be to completely replace the bottom power board in the TEOM. | ||