====== Problems and Solutions ====== | ====== Problems and Solutions ====== | ||
====== WCM ====== | ====== WCM ====== | ||
+ | |||
+ | ===== Problem (Spring 2025) ===== | ||
+ | ==== Introduction ==== | ||
+ | During previous field campaigns in which the WCM 3000 was used it had been noticed that the initial value during no water content present was a non zero value. This is still noticed in the WCM 3000 where on initial start up the probe indicated that there is water content present. This issue was normally resolved with post processing of the data from past campaigns where the non zero data would be adjusted down to zero and then further studied after that correction was applied. | ||
+ | |||
+ | ==== Cause ==== | ||
+ | In the fml.300 table that is present on the m300 computer and is used by the WCM data collection software it was noticed that the code that is supposed to correct for this non zero value wasn't included in the table. There was however correction code in there but it appeared to be from the older WCM model in which it had a compensation term that would remove this non zero value and instead keep the probe at a zero value until water content is present. | ||
+ | |||
+ | ==== Solution ==== | ||
+ | In the manual for the WCM 3000 probe there is a section in which a couple different ways are discussed as to how to get the probe to go to a zero value during the non presence of water. The way that was used in our correction of this issue was as follows; Psensedry = K1 * (Tsense - Tambient)*(Pambient * TAS)^k2. Where Psensedry is the value that is subtracted from Psensetotal which gives us Psensewet, K1 is a constant that is used which can be changed based upon testing of the probe, Tsense is the temperature of the sensor, Tambient is the ambient temperature, | ||
===== Problem (Spring 2025) ===== | ===== Problem (Spring 2025) ===== | ||
Line 54: | Line 64: | ||
Now that the long decay was removed from the graphs by removing the averaging function in the fml.300 table the PID controller should now have a more noticeable impact on the data collected. **Another thing to note which was found during the initial PID tuning was that the derivative term will also contribute to a decay since the derivative term is acting as a dampener to stop the probe from immediately going to zero after water content is stopped.** So now taking that into account when tuning the PID board it is known that the derivative term should be set fairly low so as to stop the probe from over dampening itself during data collection. The values that were found that allowed the probe to work fairly well are as follows; **(TWC terms Kp- 2000 Ki- 90 Kd- 20) (LWC terms Kp- 9500 Ki- 95 Kd- 20)**. These values were only tested during no airflow to try and calibrate the probe to be as stead as possible. This was done as a control just to factor out the unpredictability of the airflow that would normally be present over the probe and to make the tuning of the probe more easier. | Now that the long decay was removed from the graphs by removing the averaging function in the fml.300 table the PID controller should now have a more noticeable impact on the data collected. **Another thing to note which was found during the initial PID tuning was that the derivative term will also contribute to a decay since the derivative term is acting as a dampener to stop the probe from immediately going to zero after water content is stopped.** So now taking that into account when tuning the PID board it is known that the derivative term should be set fairly low so as to stop the probe from over dampening itself during data collection. The values that were found that allowed the probe to work fairly well are as follows; **(TWC terms Kp- 2000 Ki- 90 Kd- 20) (LWC terms Kp- 9500 Ki- 95 Kd- 20)**. These values were only tested during no airflow to try and calibrate the probe to be as stead as possible. This was done as a control just to factor out the unpredictability of the airflow that would normally be present over the probe and to make the tuning of the probe more easier. | ||
- | These above PID values | + | ==== Solution Two Continued... ==== |
+ | Continuing on with the second solution the PID values | ||
===== Problem (Winter 2024) ===== | ===== Problem (Winter 2024) ===== |