Line 86: Line 86:
     - Using the values you recorded and the following table, find the equation to adjust readings from the probe.     - Using the values you recorded and the following table, find the equation to adjust readings from the probe.
       - It will follow the form:       - It will follow the form:
- +        - y = mx + b 
-NaCl is 66.6, should be 75.3 +.1 +        where x = raw value 
- +        - y = corrected value
-K2SO4 is 81.5, should be +
- +
-NaCl 66.6% 73.2F, little over 23 C +
-K2SO4 81.5 72.6F, little under 23C +
 {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_132424.png?980|}} {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_132424.png?980|}}