| Line 86: | Line 86: | ||
| - Using the values you recorded and the following table, find the equation to adjust readings from the probe. | - Using the values you recorded and the following table, find the equation to adjust readings from the probe. | ||
| - It will follow the form: | - It will follow the form: | ||
| - | + | - y = mx + b | |
| - | NaCl is 66.6, should be 75.3 +- .1 | + | - where x = raw value |
| - | + | - y = corrected value | |
| - | K2SO4 is 81.5, should be | + | |
| - | + | ||
| - | NaCl 66.6% 73.2F, little over 23 C | + | |
| - | K2SO4 81.5 72.6F, little under 23C | + | |
| {{: | {{: | ||