Line 70: Line 70:
     - Keep the holes always closed when not calibrating.     - Keep the holes always closed when not calibrating.
     - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed.     - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed.
 +{{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135409.png?400|}}
     - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity.     - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity.
     - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared.     - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared.
Line 85: Line 86:
     - Using the values you recorded and the following table, find the equation to adjust readings from the probe.     - Using the values you recorded and the following table, find the equation to adjust readings from the probe.
       - It will follow the form:       - It will follow the form:
- +        - y = mx + b 
- +        - where x = raw value 
-{{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135409.png?400|}} +        y = corrected value
- +
 {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_132424.png?980|}} {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_132424.png?980|}}