atmos:citation:instruments:vaisalahmk15calibrator:home
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| atmos:citation:instruments:vaisalahmk15calibrator:home [2026/08/14 19:00] – ashley.vos | atmos:citation:instruments:vaisalahmk15calibrator:home [2026/08/17 18:32] (current) – ashley.vos | ||
|---|---|---|---|
| Line 70: | Line 70: | ||
| - Keep the holes always closed when not calibrating. | - Keep the holes always closed when not calibrating. | ||
| - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed. | - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed. | ||
| + | {{: | ||
| - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity. | - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity. | ||
| - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. | - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. | ||
| Line 85: | Line 86: | ||
| - Using the values you recorded and the following table, find the equation to adjust readings from the probe. | - Using the values you recorded and the following table, find the equation to adjust readings from the probe. | ||
| - It will follow the form: | - It will follow the form: | ||
| - | + | - y = mx + b | |
| - | + | - where x = raw value | |
| - | {{: | + | - y = corrected value |
| - | + | ||
| {{: | {{: | ||
atmos/citation/instruments/vaisalahmk15calibrator/home.1786734011.txt.gz · Last modified: 2026/08/14 19:00 by ashley.vos