Line 70: Line 70:
     - Keep the holes always closed when not calibrating.     - Keep the holes always closed when not calibrating.
     - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed.     - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed.
 +{{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135409.png?400|}}
     - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity.     - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity.
     - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared.     - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared.
   - Calibrate Instrument   - Calibrate Instrument
     - Remove the cover from the end of the thermometer and flip it around the other way.     - Remove the cover from the end of the thermometer and flip it around the other way.
-{{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135625.png?400|}}   {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135604.png?400|}}+    {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135625.png?400|}}       {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135604.png?400|}}
     - Complete the following steps twice, once for each solution. Ensure that the probe head does not get wet.     - Complete the following steps twice, once for each solution. Ensure that the probe head does not get wet.
       - Insert the thermometer into the largest hole of the calibration cover of the solution.       - Insert the thermometer into the largest hole of the calibration cover of the solution.
       - Remove the cover on the probe and insert into the calibration cover.       - Remove the cover on the probe and insert into the calibration cover.
       - Wait for the reading to stabilize, and record both the humidity reading and the temperature       - Wait for the reading to stabilize, and record both the humidity reading and the temperature
-        - This could take 10-30 minutes +        - This could take 10-30 minutes. 
-    - Repeat for the second solution.+{{:undefined:screenshot_2026-08-14_134332.png?200|}} 
 +      - Repeat for the second solution.
   - Adjust instruments   - Adjust instruments
     - Using the values you recorded and the following table, find the equation to adjust readings from the probe.     - Using the values you recorded and the following table, find the equation to adjust readings from the probe.
       - It will follow the form:       - It will follow the form:
-{{:undefined:screenshot_2026-08-14_134332.png?200|}} +        y = mx + b 
- +        where x = raw value 
-{{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_135409.png?400|}} +        y = corrected value
- +
 {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_132424.png?980|}} {{:atmos:citation:instruments:vaisalahmk15calibrator:screenshot_2026-08-14_132424.png?980|}}