| - Keep the holes always closed when not calibrating. | - Keep the holes always closed when not calibrating. | ||
| - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed. | - Write the preparation date on a sticker and use it to mark the chamber. If you use a ready-dosed salt package, use the sticker with the batch code. Mark all parts of the salt chamber (chamber, chamber lid, and transit cover) with stickers. This way, the lids/covers and different salts do not get mixed. | ||
| + | {{: | ||
| - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity. | - Allow approximately 24 hours for stabilization before use so that the salt solution reaches the equilibrium humidity. | ||
| - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. | - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. | ||
| - Calibrate Instrument | - Calibrate Instrument | ||
| - Remove the cover from the end of the thermometer and flip it around the other way. | - Remove the cover from the end of the thermometer and flip it around the other way. | ||
| + | {{: | ||
| - Complete the following steps twice, once for each solution. Ensure that the probe head does not get wet. | - Complete the following steps twice, once for each solution. Ensure that the probe head does not get wet. | ||
| - Insert the thermometer into the largest hole of the calibration cover of the solution. | - Insert the thermometer into the largest hole of the calibration cover of the solution. | ||
| - Remove the cover on the probe and insert into the calibration cover. | - Remove the cover on the probe and insert into the calibration cover. | ||
| - Wait for the reading to stabilize, and record both the humidity reading and the temperature | - Wait for the reading to stabilize, and record both the humidity reading and the temperature | ||
| - | - This could take 10-30 minutes | + | - This could take 10-30 minutes. |
| - | - Repeat for the second solution. | + | {{: |
| + | | ||
| - Adjust instruments | - Adjust instruments | ||
| - Using the values you recorded and the following table, find the equation to adjust readings from the probe. | - Using the values you recorded and the following table, find the equation to adjust readings from the probe. | ||
| - It will follow the form: | - It will follow the form: | ||
| - | {{: | + | |
| - | + | - where x = raw value | |
| - | {{: | + | - y = corrected value |
| - | {{: | + | |
| - | + | ||
| - | {{: | + | |
| {{: | {{: | ||