| Steps: | Steps: | ||
| - Prepare salt solutions | - Prepare salt solutions | ||
| - | - gg | + | - |
| - | - Depending on the frequency of use and the general operating conditions, the salt solutions | + | - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. |
| - | maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a | + | |
| - | visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the | + | |
| - | bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. | + | |
| - Wait 24 hours | - Wait 24 hours | ||
| - Calibrate Instrument | - Calibrate Instrument | ||
| - | - | ||