atmos:citation:instruments:vaisalahmk15calibrator:home
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| atmos:citation:instruments:vaisalahmk15calibrator:home [2026/07/06 17:55] – ashley.vos | atmos:citation:instruments:vaisalahmk15calibrator:home [2026/07/06 23:02] (current) – ashley.vos | ||
|---|---|---|---|
| Line 26: | Line 26: | ||
| * K2SO4 (97% RH) | * K2SO4 (97% RH) | ||
| * Calibration/ | * Calibration/ | ||
| - | * {{: | + | * {{: |
| * Certificate of Calibration/ | * Certificate of Calibration/ | ||
| - | * {{: | + | * {{: |
| Line 40: | Line 40: | ||
| Steps: | Steps: | ||
| - Prepare salt solutions | - Prepare salt solutions | ||
| + | - | ||
| + | - Depending on the frequency of use and the general operating conditions, the salt solutions maintain their characteristics for 6 - 12 months, after which they must be replaced. Perform a visual check every 2 - 3 months. There must be a minimum of 10 % of undissolved salt at the bottom of the chamber (max. 90 %). The salt must be clean, otherwise it must be reprepared. | ||
| - Wait 24 hours | - Wait 24 hours | ||
| - Calibrate Instrument | - Calibrate Instrument | ||
| + | - | ||
atmos/citation/instruments/vaisalahmk15calibrator/home.1783360534.txt.gz · Last modified: 2026/07/06 17:55 by ashley.vos